BASE
BASE-30 Automatic Solvent Extractor

Rapid solvent extraction technology is a pre-treatment technology that uses organic solvents to extract components to be measured in solid and semi-solid samples at higher temperatures (50-200°C) and higher pressures.


Ø rapid sample pretreatment device specially designed for solid and semi-solid samples based on the above principles.

Ø suitable for existing chromatography, chromatography-mass spectrometry Analytical instrument sample preprocessing.

Ø Solvent loading, heating, pressurization, extraction, rinsing, filtration, purification, collection, and nitrogen blowing are all completed automatically.

Ø Three samples can be extracted simultaneously at one time, and each round can automatically complete the extraction of up to 30 samples.

Ø The volume of each extraction cell can be selected from 22ml, 34ml, and 66ml.

Ø The entire extraction process can be unattended. If any operation is abnormal, it can alarm automatically and stop working if abnormal


Application

Rapid extraction of target components from solid or semi-solid samples in soil, environmental, chemical, pharmaceutical, electronic products, agricultural and food products and some natural product samples. 


Instrument performance indicators

Ø Can be placed in a fume hood for experimental operations

Ø Overall dimensions: 1100*670*730mm

Ø Full circumferential heating furnace design, temperature control range: room temperature ~ 200°C;

Ø Extraction tank volume: 34ml, 66ml,

Ø Extraction time: single extraction is about 20 minutes,

Ø The extract can be automatically filtered

Ø It can automatically extract 26 samples at a time

Ø Solvent controller (optional): According to the preset mixed solvent ratio, it can automatically complete the mixing ratio of up to 4 solvents


Standard:

EPA  3545A-1996 OCP, OPP,BNA,TPH,PCDD


Last:No Data